Home
SolutionsNews & eventsService & supportEducationAbout FOSSJobsPressContactLogin
Home » Solutions » Products » BactoScan™ FC » Technical specifications
Home
Solutions
Products
BactoScan™ FC
BactoScan™ FC Semiautomatic
Benefits and features
Applications and parameters
Payback on investment
Technical specifications
Technology

Next step
Request further info
Newsletter sign up

Technical specifications

Application data
Capacity

Models 50L, 50H:
50 samples per hour

Models 100L, 100H:
100 samples per hour

Models 150L, 150H:
150 samples per hour

Analysis time

9 minutes

Sample intake

4.5 ml

Sample temperature

2 - 42ºC

Sample quality

Raw milk of normal composition and good quality from cows. Un-preserved or preserved with azidiol.

Performance data
Typical measuring range

L Models:
From 100,000 to minimum 20 million IBC/ml (approx. 30,000 to 10 million CFU/ml)

H Models:
From 5,000 to minimum 20 million IBC/ml (approx. 1,500 to 10 million CFU/ml)

Repeatability

 

Range (IBC/µl): Sr (log-units): Typical Sr (log-units)
10-50 0.07 0.06
51-200 0.05 0.04
200 0.04 0.02
Entire range 0.05

Reproducibility (between instruments)

 

Range (IBC/µl): SR (log-units): Typical SR (log-units)
10-50 0.11 0.08
51-200 0.07 0.06
>200 0.06 0.04

 

Carry-over effect

< 0.5% (uncompensated)

Working factor

300

Accuracy

Typical Sy,x < 0.25 log units in the entire measuring range.

Reference method

Standard Plate Count (IDFStandard 100B: 1991). 


Related products

SolutionsNews & eventsService & supportEducationAbout FOSSJobsPressContactLogin
© FOSS 1956 - 2010 All rights reserved   Legal notice   About this site   Feedback   Site map